Quality of chest compressions by healthcare professionals using real-time audiovisual feedback during in-hospital cardiopulmonary resuscitation

European Journal of Cardiovascular Nursing

1 June 2017
Organised by: Logo
ESC Journals

Abstract

AbstractIntroduction

A high quality of chest compressions, e.g. sufficient depth (5–6 cm) and rate (100–120 per min), has been associated with survival. The patient’s underlay affects chest compression depth. Depth and rate can be assessed by feedback systems to guide rescuers during cardiopulmonary resuscitation.

Aim

The purpose of this study was to describe the quality of chest compressions by healthcare professionals using real-time audiovisual feedback during in-hospital cardiopulmonary resuscitation.

Method

An observational descriptive study was performed including 63 cardiac arrest events with a resuscitation attempt. Data files were recorded by Zoll AED Pro, and reviewed by RescueNet Code Review software. The events were analysed according to depth, rate, quality of chest compressions and underlay.

Results

Across events, 12.7% (median) of the compressions had a depth of 5–6 cm. Compression depth of >6 cm was measured in 70.1% (median). The underlay could be identified from the electronic patient records in 54 events. The median compression depth was 4.5 cm (floor) and 6.7 cm (mattress). Across events, 57.5% (median) of the compressions were performed with a median frequency of 100–120 compressions/min and the most common problem was a compression rate of <100 (median=22.3%).

Conclusions

Chest compression quality was poor according to the feedback system. However, the distribution of compression depth with regard to underlay points towards overestimation of depth when treating patients on a mattress. Audiovisual feedback devices ought to be further developed. Healthcare professionals need to be aware of the strengths and weaknesses of their devices.