Cryoablation vs. radiofrequency ablation of the atrioventricular node in patients with His-bundle pacing
EP Europace Journal

Abstract
Radiofrequency ablation (RFA) of the atrioventricular node (AVN) with His-bundle pacing (HBP) can cause rise in capture thresholds. Cryoablation (CRYO) may offer reversibility in case of threshold rise but has never been tested for AVN ablation in this setting. Our aim was to compare procedural characteristics and outcome of CRYO compared with RFA for AVN ablation in patients with HBP.
Forty-four patients with HBP underwent AVN ablation for an ‘ablate and pace’ indication. Cryoablation was performed in the first 22 patients and RFA in the following 22 patients. Procedural characteristics, success rates, and change in His capture thresholds were compared between groups. Distance from the ablation site to the His lead was measured using biplane fluoroscopy. Acute success was 100% with both strategies. Median procedural duration was significantly longer for CRYO {50 [interquartile range (IQR) 38–63] min} compared with RFA [36 (IQR, 30–41) min;
Cryoablation does not offer any advantage over RFA for AVN ablation in patients with HBP and tended to require more redo procedures. If possible, a distance of ≥6 mm should be maintained from the His lead tip to avoid a rise in capture thresholds.