European experience with a first totally leadless cardiac resynchronization therapy pacemaker system
EP Europace Journal

Abstract
Totally leadless cardiac resynchronization therapy (CRT) can be delivered with a combination of Micra and WiSE-CRT systems. We describe the technical feasibility and first insights into the safety and efficacy of this combination in European experience.
Patients enrolled had indication for both Micra and WiSE-CRT systems because of heart failure related to high burden of pacing by a Micra necessitating system upgrade or inability to implant a conventional CRT system because of infectious or anatomical conditions. The endpoints of the study were technical success of WiSE-CRT implantation with right ventricle-synchonized CRT delivery, acute QRS duration reduction, and freedom from procedure-related major adverse events. All eight WiSE-CRT devices were able to detect the Micra pacing output and to be trained to deliver synchronous LV endocardial pacing. Acute QRS reduction following WiSE-CRT implantation was observed in all eight patients (mean QRS 204.38 ± 30.26 vs. 137.5 ± 24.75 mS,
The Micra and the WiSE-CRT systems can successfully operate together to deliver total leadless CRT to a patient. Moreover, the WiSE-CRT system provides the only means to upgrade the large population of Micra patients to CRT capability without replacing the Micra. The range of application of this combination could broaden in the future with the upcoming developments of leadless cardiac pacing.
Contributors

Adrien Carabelli
Author

Mariem Jabeur
Author

Peggy Jacon
Author

Christopher Aldo Rinaldi
Author

Christophe Leclercq
Author

Giovanni Rovaris
Author

Martin Arnold
Author

Sandrine Venier
Author

Petr Neuzil
Author