Acute oesophageal safety of high-power short duration with 50 W for atrial fibrillation ablation
EP Europace Journal

Abstract
Pulmonary vein isolation (PVI) using radiofrequency (RF) ablation is an effective treatment option for patients with atrial fibrillation (AF). This study aims to investigate the safety of high-power short duration (HPSD) with emphasis on oesophageal lesions after PVI.
Consecutive patients undergoing AF ablation with HPSD (50 W; ablation index (AI)-guided; target AI 350 for posterior wall ablation, AI 450 for anterior wall ablation) using the ThermoCool SmartTouch SF catheter were included. Patients underwent post-ablation oesophageal endoscopy to detect and categorize thermal oesophageal injury (EDEL). Occurrence and risk factors of oesophageal lesions and perforating complications were analysed. A total of 1033 patients underwent AF ablation with HPSD. Of them, 953 patients (67.6 ± 9.6 years; 58% male; 43% paroxysmal AF; 68% first PVI) underwent post-procedural oesophageal endoscopy and were included in further analyses. Median procedure time was 82.8 ± 24.4 min with ablation times of 16.1 ± 9.2 min. Thermal oesophageal injury was detected in 58 patients (6%) (
Thermal oesophageal lesions occur in 6% of HPSD AF ablations. The risk for development of perforating complications seems to be low. Incidence of atrio-oesophageal fistula (0.1%) is comparable to other reported series about RF ablation approaches.
Contributors

Artur Berkovitz
Author

Karin Nentwich
Author

Elena Ene
Author

Kai Sonne
Author

Gelu Simu
Author

Ivalyo Chakarov
Author

Sebastian Barth
Author

Christian Waechter
Author

Michael Behnes
Author

Ibrahim Akin
Author

Thomas Deneke
Author
You may be interested in