Long-term outcomes after transcatheter aortic valve implantation in failed bioprosthetic valves

European Heart Journal

27 June 2020
Organised by: Logo
ESC Journals VALVULAR, MYOCARDIAL, PERICARDIAL, PULMONARY, CONGENITAL HEART DISEASE Valvular Heart Disease

Abstract

AbstractAims

Due to bioprosthetic valve degeneration, aortic valve-in-valve (ViV) procedures are increasingly performed. There are no data on long-term outcomes after aortic ViV. Our aim was to perform a large-scale assessment of long-term survival and reintervention after aortic ViV.

Methods and results

A total of 1006 aortic ViV procedures performed more than 5 years ago [mean age 77.7 ± 9.7 years; 58.8% male; median STS-PROM score 7.3% (4.2–12.0)] were included in the analysis. Patients were treated with Medtronic self-expandable valves (CoreValve/Evolut, Medtronic Inc., Minneapolis, MN, USA) (n = 523, 52.0%), Edwards balloon-expandable valves (EBEV, SAPIEN/SAPIEN XT/SAPIEN 3, Edwards Lifesciences, Irvine, CA, USA) (n = 435, 43.2%), and other devices (n = 48, 4.8%). Survival was lower at 8 years in patients with small-failed bioprostheses [internal diameter (ID) ≤ 20 mm] compared with those with large-failed bioprostheses (ID > 20 mm) (33.2% vs. 40.5%, P = 0.01). Independent correlates for mortality included smaller-failed bioprosthetic valves [hazard ratio (HR) 1.07 (95% confidence interval (CI) 1.02–1.13)], age [HR 1.21 (95% CI 1.01–1.45)], and non-transfemoral access [HR 1.43 (95% CI 1.11–1.84)]. There were 40 reinterventions after ViV. Independent correlates for all-cause reintervention included pre-existing severe prosthesis–patient mismatch [subhazard ratio (SHR) 4.34 (95% CI 1.31–14.39)], device malposition [SHR 3.75 (95% CI 1.36–10.35)], EBEV [SHR 3.34 (95% CI 1.26–8.85)], and age [SHR 0.59 (95% CI 0.44–0.78)].

Conclusions

The size of the original failed valve may influence long-term mortality, and the type of the transcatheter valve may influence the need for reintervention after aortic ViV.

Contributors

Luca Testa
Luca Testa

Author

IRCCS San Donato Polyclinic San Donato Milanese , Italy

Antonio Landi
Antonio Landi

Author

Cardiocentro Ticino Institute Lugano , Switzerland

Stefano Salizzoni
Stefano Salizzoni

Author

A.O.U. Citta Della Salute E Della Scienza Di Torino Molinette Turin , Italy

Danny Dvir
Danny Dvir

Author

Shaare Zedek Medical Center Jerusalem , Israel

ESC 365 is supported by