Echocardiographic determinants of successful balloon dilation in pulmonary atresia with intact ventricular septum
European Heart Journal - Cardiovascular Imaging

Abstract
Pulmonary atresia with intact ventricular septum (PA–IVS) is a complex congenital heart malformation with multitude therapeutic approaches. Recently, balloon valvotomy has been used as an alternative to primary surgery. This study aimed to identify echocardiographic markers of balloon dilation success in PA–IVS.
The echocardiograms of 26 patients diagnosed with PA–IVS who underwent primary pulmonary balloon valvotomy were reviewed. Tricuspid annulus
Successful balloon dilation in membranous type PA–IVS can be predicted by a scoring system using RV/LV diameter ratio and RV/LV length ratio.