Comparison of three-dimensional echocardiographic software packages for assessment of left ventricular mechanical dyssynchrony and prediction of response to cardiac resynchronization therapy
European Heart Journal - Cardiovascular Imaging

Abstract
We directly compared TomTec and QLAB software packages for the three-dimensional echocardiographic (3DE) assessment of left ventricular (LV) dyssynchrony including their ability to predict response to cardiac resynchronization therapy (CRT) in patients with ischaemic and non-ischaemic cardiomyopathy.
A total of 140 heart failure patients with the LVEF ≤35% and 60 healthy volunteers underwent 3DE. A subgroup of 60 patients underwent CRT and were evaluated before and 6–12 months after implantation. The systolic dyssynchrony index (SDI) was derived from the dispersion of time to minimum regional volume for all 16 LV segments and measured with both software packages and compared using Pearson's correlation and Bland–Altman analysis. Measurements of SDI were significantly higher using TomTec compared with QLAB in both patients (10.9 ± 3.8 vs. 9.7 ± 3.9,
Different 3DE software packages for the assessment of mechanical dyssynchrony should not be used interchangeably until better software standardization is achieved. Dyssynchrony assessment with 3DE for the prediction of response to CRT seems particularly useful in patients with non-ischaemic cardiomyopathy.
Contributors

Sebastiaan A. Kleijn
Author

Karin de Boer
Author

Yasser A. Abd El-Hady
Author

Khalid A. Sorour
Author

Hossam I. Kandil
Author

Albert C. van Rossum
Author

Otto Kamp
Author