Large-field intravascular ultrasound for annular sizing and predicting paravalvular regurgitation during TAVR: comparisons with multidetector computed tomography and transoesophageal echocardiography
European Heart Journal - Cardiovascular Imaging

Abstract
The use of contrast media with multidetector computed tomography (MDCT) may induce acute kidney injury in patients with renal failure undergoing transcatheter aortic valve replacement (TAVR). We investigated the role of large-field intravascular ultrasound (IVUS) vs. MDCT and two-dimensional transoesophageal echocardiography (2D-TEE) for annular sizing and predicting paravalvular regurgitation (PVR) during TAVR.
The aortic annulus was measured by large-field IVUS and 2D-TEE, and compared with MDCT in 50 patients undergoing TAVR. The IVUS and MDCT annular areas and diameters were not significantly different (446 ± 87 mm2 and 23.8 ± 84 mm vs. 466 ± 84 mm2 and 24 ± 2.1 mm, respectively;
The aortic annular measurements and predicting PVR by large field IVUS were not significantly different from those of MDCT, but were greater than those of TEE. Large filed IVUS can be reliably used in lieu of MDCT for annular sizing in patients with aortic stenosis and renal failure or suboptimal MDCT images.