Insertable cardiac monitors in the diagnosis of syncope and the detection of atrial fibrillation: A systematic review and meta-analysis
European Journal of Preventive Cardiology

Abstract
Insertable or implantable cardiac monitors (ICMs) continuously monitor the heart rhythm and record irregularities over 3 years, enabling the diagnosis of infrequent rhythm abnormalities associated with syncope and stroke. The enhanced recognition capabilities of recent ICM models are able to accurately detect atrial fibrillation (AF) and have led to new applications of ICMs for the detection and monitoring of AF.
Based on a systematic literature search, two indications were identified for ICMs for which considerable evidence, including randomized studies, exists: diagnosing the underlying cardiac cause of unexplained recurrent syncope and detecting AF in patients after cryptogenic stroke (CS). Three randomized controlled trials (RCTs) were identified that compared the effectiveness of ICMs in diagnosing patients with unexplained syncope (
Prolonged monitoring with ICMs is an effective tool for diagnosing the underlying cardiac cause of unexplained syncope and for detecting AF in patients with CS. In all RCTs, ICMs have a superior diagnostic yield compared to conventional monitoring.