Assessment of cardiac image quality from single-probe handheld ultrasound devices

European Heart Journal - Imaging Methods and Practice

6 October 2026
Organised by: Logo
ESC Journals IMAGING Echocardiography VALVULAR, MYOCARDIAL, PERICARDIAL, PULMONARY, CONGENITAL HEART DISEASE Valvular Heart Disease

Abstract

AbstractAims

This study aims to assess quality of cardiac ultrasound imaging from commercially available single-probe handheld devices.

Methods and results

Cardiac imaging from single-probe handheld ultrasound devices was prospectively acquired by registered cardiac sonographers. Videos were graded on a scale of 1 (uninterpretable) to 5 (diagnostic-quality imaging) by 2 sonographers. The average image quality score for each cine loop was analysed. Interpretable imaging was defined as an average IQ score ≥ 2. Good image quality was defined as an average IQ score ≥ 4. Imaging from single-probe devices representing 518 videos from 206 patients was assessed. The mean age was 72 (SD 11.1) years. Agreement on image quality was moderate (weighted kappa 0.50, 95% CI 0.45–0.55). The correlation between sonographer quality grades was 0.60. The median average IQ score was 2.5 (1.5). 454 (88%) of the videos were interpretable (average IQ ≥ 2). At least one interpretable video was obtained for 187 (91%) patients. 62/518 (12%) of videos had good image quality (average score ≥ 4). There was a trend towards lower image quality with increasing BMI categories [16–25, median 3.0 (2.0, 3.5), 25–29, median 3.0 (2.0, 3.5), > 29 median 2.5 (2.0, 3.0), P < 0.001].

Conclusion

Basic interpretable imaging from single-probe handheld ultrasound devices can be obtained for most patients under idealized conditions, suggesting possible use for enhanced cardiac disease detection. Since imaging from a significant minority of patients is uninterpretable and lower image quality is associated with higher BMI, rigorous assessment of image quality is essential prior to image interpretation.