Impella preserves haemodynamics with adequate stressed blood volume and normal pulmonary vascular resistance in a goat model of ventricular fibrillation
EHJ - Open

Abstract
Ventricular fibrillation (VF) is a frequent and life-threatening complication of acute myocardial infarction. Although the Impella device provides circulatory support solely to the left ventricle (LV), emerging reports have suggested that it may partially maintain haemodynamics even during VF. This study sought to examine this phenomenon under controlled experimental conditions.
In eight Saanen goats, we recorded right atrial pressure (RAP), left atrial pressure (LAP), arterial pressure (AP), LV pressure, pulmonary artery flow (PAF), carotid artery flow (CAF), and coronary flow (CoF). We inserted an Impella CP and induced VF using direct current. In Protocol 1, we changed the stressed blood volume (SBV) by blood withdrawal/infusion or pulmonary vascular resistance (PVR) by pulmonary microembolisation and measured the haemodynamics during VF with Impella P4 support (
Impella preserved systemic, cerebral, and coronary blood flow with adequate SBV and normal PVR conditions in a goat model of VF.