Machine learning-enabled screening for aortic stenosis with handheld ultrasound

European Heart Journal - Imaging Methods and Practice

8 May 2025
Organised by: Logo
ESC Journals IMAGING Echocardiography VALVULAR, MYOCARDIAL, PERICARDIAL, PULMONARY, CONGENITAL HEART DISEASE Valvular Heart Disease

Abstract

AbstractAims

Neural network classifiers can detect aortic stenosis (AS) using limited cardiac ultrasound images. While networks perform very well using cart-based imaging, they have never been tested or fine-tuned for use with focused cardiac ultrasound (FoCUS) acquisitions obtained on handheld ultrasound devices.

Methods and results

Prospective study performed at Tufts Medical Center. All patients ≥65 years of age referred for clinically indicated transthoracic echocardigraphy (TTE) were eligible for inclusion. Parasternal long axis and parasternal short axis imaging was acquired using a commercially available handheld ultrasound device. Our cart-based AS classifier (trained on ∼10 000 images) was tested on FoCUS imaging from 160 patients. The median age was 74 (inter-quartile range 69–80) years, 50% of patients were women. Thirty patients (18.8%) had some degree of AS. The area under the received operator curve (AUROC) of the cart-based model for detecting AS was 0.87 (95% CI 0.75–0.99) on the FoCUS test set. Last-layer fine-tuning on handheld data established a classifier with AUROC of 0.94 (0.91–0.97). AUROC during temporal external validation was 0.97 (95% CI 0.89–1.0). When performance of the fine-tuned AS classifier was modelled on potential screening environments (2 and 10% AS prevalence), the positive predictive value ranged from 0.72 (0.69–0.76) to 0.88 (0.81–0.97) and negative predictive value ranged from 0.94 (0.94–0.94) to 0.99 (0.99–0.99) respectively.

Conclusion

Our cart-based machine-learning model for AS showed a drop in performance when tested on handheld ultrasound imaging collected by sonographers. Fine-tuning the AS classifier improved performance and demonstrates potential as a novel approach to detecting AS through automated interpretation of handheld imaging.