Leadless vs. transvenous single-chamber ventricular pacing in the Micra CED study: 2-year follow-up
European Heart Journal

Abstract
Clinical trials have demonstrated the safety and efficacy of the Micra leadless VVI pacemaker; however, longer-term outcomes in a large, real-world population with a contemporaneous comparison to transvenous VVI pacemakers have not been examined. We compared reinterventions, chronic complications, and all-cause mortality at 2 years between leadless VVI and transvenous VVI implanted patients.
The Micra Coverage with Evidence Development study is a continuously enrolling, observational, cohort study of leadless VVI pacemakers in the US Medicare fee-for-service population. Patients implanted with a leadless VVI pacemaker between March 9, 2017, and December 31, 2018, were identified using Medicare claims data linked to manufacturer device registration data (
In a real-world study of US Medicare patients, the Micra leadless VVI pacemaker was associated with a 38% lower adjusted rate of reinterventions and a 31% lower adjusted rate of chronic complications compared with transvenous VVI pacing. There was no difference in adjusted all-cause mortality at 2 years.