Mid-term performance of His bundle pacing and usefulness of backup leads

EP Europace Journal

14 June 2024
Organised by: Logo
ESC Journals ARRHYTHMIAS AND DEVICE THERAPY Device Therapy

Abstract

Abstract

Ventricular backup leads may be considered in selected patients with His bundle pacing (HBP), but it remains unknown to what extent this is useful. A total of 184 HBP patients were studied. At last follow-up, 147 (79.9%) patients retained His bundle capture at programmed output. His bundle pacing lead revision was performed in 5/36 (13.9%) patients without a backup lead and in 3/148 (2.0%) patients with a backup lead (P = 0.008). One patient without a backup lead had syncope due to atrial oversensing. Thus, implantation of ventricular backup leads may avoid lead revision and adverse events in selected HBP patients.

Contributors

Haran Burri
Haran Burri

Author

University hospitals of Geneva Geneva , Switzerland