Evaluation of a novel cuffless photoplethysmography-based wristband for measuring blood pressure according to the regulatory standards

European Heart Journal - Digital Health

8 February 2024
Organised by: Logo
ESC Journals CARDIOVASCULAR NURSING AND ALLIED PROFESSIONS CORONARY ARTERY DISEASE, ACUTE CORONARY SYNDROMES, ACUTE CARDIAC CARE HYPERTENSION PREVENTIVE CARDIOLOGY Risk Factors and Prevention

Abstract

AbstractAims

Elevated blood pressure (BP) is a key risk factor in cardiovascular diseases. However, obtaining reliable and reproducible BP remains a challenge. This study, therefore, aimed to evaluate a novel cuffless wristband, based on photoplethysmography (PPG), for continuous BP monitoring.

Methods and results

Predictions by a PPG-guided algorithm were compared to arterial BP measurements (in the sub-clavian artery), obtained during cardiac catheterization. Eligible patients were included and screened based on AAMI/European Society of Hypertension (ESH)/ISO Universal Standard requirements. The machine learning-based BP algorithm required three cuff-based initialization measurements in combination with ∼100 features (signal-derived and patient demographic-based). Ninety-seven patients and 420 samples were included. Mean age, weight, and height were 67.1 years (SD 11.1), 83.4 kg (SD 16.1), and 174 cm (SD 10), respectively. Systolic BP was ≤100 mmHg in 48 samples (11%) and ≥160 mmHg in 106 samples (25%). Diastolic BP was ≤70 mmHg in 222 samples (53%) and ≥85 mmHg in 99 samples (24%). The algorithm showed mean errors of ±3.7 mmHg (SD 4.4 mmHg) and ±2.5 mmHg (SD 3.7 mmHg) for systolic and diastolic BP, respectively. Similar results were observed across all genders and skin colours (Fitzpatrick I-VI).

Conclusion

This study provides initial evidence for the accuracy of a PPG-based BP algorithm in combination with a cuffless wristband across a range of BP distributions. This research complies with the AAMI/ESH/ISO Universal Standard, however, further research is required to evaluate the algorithms performance in light of the remaining European Society of Hypertension recommendations.

Clinical trial registration

www.clinicaltrials.gov, NCT05566886.

Contributors

Eelko Ronner
Eelko Ronner

Author

Reinier de Graaf Hospital Delft , Netherlands (The)

ESC 365 is supported by