Electrical characteristics of pacing different portions of the His bundle in bradycardia patients
EP Europace Journal

Abstract
His-bundle pacing (HBP) can be achieved in either atrial-side HBP (aHBP) or ventricular-side HBP (vHBP). The study compared the pacing parameters and electrophysiological characteristics between aHBP and vHBP in bradycardia patients.
Fifty patients undergoing HBP implantation assisted by visualization of the tricuspid valvular annulus (TVA) were enrolled. The HBP lead position was identified by TVA angiography. Twenty-five patients were assigned to undergo aHBP and compared with 25 patients who underwent vHBP primarily in a prospective and randomized fashion. Pacing parameters and echocardiography were routinely assessed at implant and 3-month follow-up. His-bundle pacing was successfully performed in 45 patients (90% success rate with 44.4% aHBP and 55.6% vHBP). The capture threshold was lower in vHBP than aHBP at implant (vHBP: 1.1 ± 0.5 vs. aHBP: 1.4 ± 0.4 V/1.0 ms,
This study demonstrates that vHBP features a low and stable pacing capture threshold and high R-wave amplitude, suggesting better pacing mode management and battery longevity can be achieved by HBP in the ventricular side.