Real-time intracardiac echocardiography validation of saline-enhanced radiofrequency needle-tip ablation: lesion characteristics and gross pathology correlation
EP Europace Journal

Abstract
With the implementation of saline-enhanced radiofrequency (SERF) needle-tip ablation, real-time validation of lesion formation is needed for the controllable creation of transmural lesions. The aim of the study was to analyse the ability of two-dimensional intracardiac echocardiography (2D-ICE) to guide and validate SERF ablation in real-time.
Fifty-six SERF energy deliveries at left ventricular sites of 11 dogs guided by 2D-ICE were analysed (power: 15–50 W; time: 25–120 s; irrigation saline: 60°C with 10 mL/min flow rate). Catheter tip/tissue orientation and lesion formation could be well detected by 2D-ICE in 49 (87.5%) energy deliveries. Gross pathology analysis confirmed excellent 2D-ICE lesion localization, the ability to detect transmural lesions (70% sensitivity, 47% specificity) and positive correlation between 2D-ICE and the corresponding gross pathology measurements of ‘maximal lesion depth’; (repeated measures correlation:
With excellent lesion localization and a 70% detection rate of transmural lesions, 2D-ICE is well suited to validate SERF ablation lesion formation in real-time. The catheter tip/tissue angle impacts the lesion formation and through perpendicular catheter positioning, deeper intramural areas of the myocardium can be reached.
Contributors

Songyun Wang
Author

Atsushi Suzuki
Author

Kimitake Imamura
Author

H Immo Lehmann
Author

Kay D Parker
Author

Laura K Newman
Author

Kristi H Monahan
Author

Maryam E Rettmann
Author

Michael G Curley
Author

Douglas L Packer
Author
Intermountain Cardiovascular Research Institute Salt Lake City , United States of America