Impact of closed loop stimulation on prognostic cardiopulmonary variables in patients with chronic heart failure and severe chronotropic incompetence: a pilot, randomized, crossover study
EP Europace Journal

Abstract
Clinical effects of rate-adaptive pacing in heart failure patients with chronotropic incompetence (CI) undergoing cardiac resynchronization therapy (CRT) remain unclear. Closed loop stimulation (CLS) is a new rate-adaptive sensor in CRT devices. We evaluated the effectiveness of CLS in CRT patients with severe CI, focusing primarily on key prognostic variables assessed by cardiopulmonary exercise (CPX) testing.
In the randomized, crossover, multicentre BIO|CREATE study, 20 CRT patients with severe CI and NYHA Class II/III (60%/40%) were randomized 1:1 to the sequence DDD-40 mode to DDD-CLS mode, or the sequence DDD-CLS mode to DDD-40 mode (1 month in each mode). Patients underwent symptom-limited treadmill-based CPX test in each mode. An improvement (decrease) of the ventilatory efficiency (VE) slope of ≥5% during CLS was regarded as positive response to CLS. Seventeen patients with full data sets had a mean intra-individual VE slope change of −1.8 ± 3.0 (−4.1%) with CLS (
The data of the present pilot study are compatible with the notion that CLS activation may improve VE slope in CRT patients with severe CI and less advanced heart failure. Further research is needed to determine the long-term clinical outcomes of CLS.
Contributors

Joachim Proff
Author

Béla Merkely
Author

Roland Papp
Author

Corinna Lenz
Author

Peter Nordbeck
Author

Christian Butter
Author

Juergen Meyerhoefer
Author

Michael Doering
Author

Dean J MacCarter
Author

Katharina Ingel
Author

Thomas Thouet
Author

Mattias J Roser
Author