ESC Professional Members, Association Members (Ivory & above) benefit from year-round access to all the resources from their respective Association, and to all content from previous years. Fellows of the ESC (FESC), and professionals in training or under 40 years old, who subscribed to a Young Combined Membership package benefit from access to all ESC 365 content from all events, all editions, all year long. Find out more about ESC Memberships here.
An automatic method for myocardial T2* curve fitting in thalassemia patients with severe iron overload
Topic : e-cardiology/digital health. public health. health economics. research methodology
Sub-topic : Cardiovascular Image Processing
Session type : Abstract Sessions
Authors : A Meloni (Pisa,IT), V Positano (Pisa,IT), L Gulino (Pisa,IT), L Landini (Pisa,IT), F Santarelli (Pisa,IT), V Santamaria (Vibo Valenzia,IT), G Restaino (Campobasso,IT), S Renne (Lamezia Terme,IT), M Lombardi (Pisa,IT), A Pepe (Pisa,IT)
A. Meloni1
,
V. Positano1
,
L. Gulino1
,
L. Landini1
,
F. Santarelli1
,
V. Santamaria2
,
G. Restaino3
,
S. Renne4
,
M. Lombardi1
,
A. Pepe1
,
1CMR Unit, Fondazione G.Monasterio CNR-Regione Toscana and Institute of Clinical Physiology - Pisa - Italy
,
2Immunologia e Centro Trasfusionale - ASP - Vibo Valenzia - Italy
,
3Catholic University of the Sacred Heart, Department of Radiology - Campobasso - Italy
,
4Giovanni Paolo II Hospital - UTIC, Department of Cardiology - Lamezia Terme - Italy
,
Purpose: Myocardial iron overload assessment by multislice multiecho T2* technique is used in the clinical management of thalassemia major (TM) patients. Signal decay curves are extracted from the 16 left ventricular (LV) segments and the fitting of these curves to a mono-exponential model provides the corresponding T2* values. In patients with severe cardiac iron overload, where signal will decay quickly becoming comparable to image noise, manual truncation of signal decay curves excluding later echo times (TEs) is adopted. In this study an automatic truncation method avoiding the variability associated with the manual selection of the truncation point is introduced and validated.
Methods: Twenty patients (13 males, age 33±7 years) enrolled in the MIOT Network and diagnosed for severe iron overload (T2*<10 ms) were considered. Using a previously validated software the segmental T2* values were evaluated by the standard methodology (i.e. manual truncation).
Images were independently analysed by the developed automated approach. The percentage fitting error (e) was computed as the root mean square error between the signal decay curve and the mono-exponential model normalized to the mean value of the signal.
If e was > 5%, the algorithm cut-off the last TE and performed again the fitting. The procedure was iterated until the error become <5% or the number of TEs become equal to three. To assess the inter-operator variability, the dataset was processed by a second operator.
Results: The Coefficient of Variability (CoV) for inter-observer variability was 6.82±4.01%. The CoV between automated and manual analysis was 6.15±3.92%, not significantly different from inter-observer variability (P=0.332). No significant difference was detected between mid-septum and global T2* values evaluated with manual and automated procedure (P=0.26 and P=0.91, respectively). The mean fitting error was not significantly different in manual and automated analysis (4.10±2.11 vs. 4.52±2.12, P=0.53). In segmental analysis, no significant differences were found between manual and automatic procedure (P>0.01 for all segments).
Conclusions: Truncation of signal decay curve needed to compensate for low signal in later echoes in patients with severe iron overload can be effectively automatized avoiding operator induced variability.
ESC Professional Members, Association Members (Ivory & above) benefit from year-round access to all the resources from their respective Association, and to all content from previous years. Fellows of the ESC (FESC), and professionals in training or under 40 years old, who subscribed to a Young Combined Membership package benefit from access to all ESC 365 content from all events, all editions, all year long. Find out more about ESC Memberships here.