Pulmonary vein isolation by different setting of high-power short-duration radiofrequency application: feasibility, short term efficacy and safety in patients with atrial fibrillation

Session Managing electrical complications after Transcatheter Aortic Valve Implantation and other percutaneous arrhythmia procedures

Speaker Jose Luis Merino

Event : ESC Congress 2019

  • Topic : arrhythmias and device therapy
  • Sub-topic : Catheter Ablation of Arrhythmias
  • Session type : Rapid Fire Abstracts

Authors : S Castrejon Castrejon (Madrid,ES), M Martinez Cossiani (Madrid,ES), M Ortega Molina (Madrid,ES), C Escobar Cervantes (Madrid,ES), C Froilan Torres (Madrid,ES), N Gonzalo Bada (Madrid,ES), M Diaz De La Torre (Madrid,ES), JM Suarez Parga (Madrid,ES), M Batlle (Madrid,ES), R Gonzalez Davia (Parla,ES), JL Lopez Sendon (Madrid,ES), JL Merino (Madrid,ES)

Authors:
S. Castrejon Castrejon1 , M. Martinez Cossiani1 , M. Ortega Molina1 , C. Escobar Cervantes1 , C. Froilan Torres2 , N. Gonzalo Bada2 , M. Diaz De La Torre2 , J.M. Suarez Parga2 , M. Batlle3 , R. Gonzalez Davia4 , J.L. Lopez Sendon5 , J.L. Merino1 , 1Robotic Cardiac Electrophysiology Unit, University Hospital La Paz - Madrid - Spain , 2University Hospital La Paz, Gastroenterology Acute Unit - Madrid - Spain , 3University Hospital del Henares, Service of Cardiology - Madrid - Spain , 4University Hospital Infanta Cristina, Cardiology - Parla - Spain , 5University Hospital La Paz, Department of Cardiology - Madrid - Spain ,

Citation:
European Heart Journal ( 2019 ) 40 ( Supplement ), 3736

Aims: Pulmonary vein (PV) isolation (PVI) by point-by-point radiofrequency application (PPRF) results in longer procedures than cryoballoon ablation. In addition, it is associated with more esophageal lesions. The aim of this study was to evaluate the feasibility and safety of PVI by high power short duration (HPSD) PPRF in patients with atrial fibrillation.

Methods: PPRF around the PVs was done in 125 consecutive patients distributed in two chronologically successive groups. Conventional PPRF with 30W for ≤30 s under luminal esophageal temperature monitoring was performed in the first 47 patients (Group 1). 68 patients were enrolled in the HPSD (Group 2). Power was set to 50 W and delivered to reach a predefined lesion index value (LSI ≥5 or Abl-I ≥350) in the first 18 patients (Group 2A). 30 patients underwent PPRF with 60W for 7–10 s (Group 2B) and the last 30 patients underwent PPRF with 70W for 9 s (Group 2C). Esophageal endoscopy was performed after ablation in all patients.

Results: PVI of all targeted veins was achieved in 96% and 100% of patients of groups 1 and 2 respectively (p=0.6). Total RF time was 30 [27–43], 25 [20–29], 16 [14–20] and 14 [11–16] min in groups 1, 2A and 2B and 2C respectively (p<0.001). RF was delivered for 12 [9–17] s vs 9 [8–9]s vs 9 [8–9] per application in groups 2 A, 2B and 2C respectively (p<0.001). Total number of RFa to completely isolate all PV was 105 [90–126] in group 2A, 113 [90–135] in group 2B and 94 [79–112] in group 2C (p=0.12).

First-pass PVI was achieved in 56%, 57% and 85% of left PV (p=0.038) and in 56%, 60% and 82% of right PV (p=0.13) in groups 50W, 60W and 70W respectively. The carina was the most frequent location of persistent conduction when first-pass failed. Reconnections occurred in 6%, 3% and 11% of left PV (p=0.6) and in 6%, 7% and 4% (p=0.63) of right PV in groups 50W, 60W and 70W respectively. Adenosine test was systematically used in groups 60W and 70W: the incidence of dormant conduction was 23% and 22% (p=0.9) in left PV and 20% and 22% (p=0.8) in right PV respectively.

The incidence of esophageal lesions was 28% in Group 1, 2% in Group 2A, and 0% in groups 2B and 2C (p<0.001). No other intraprocedural complications occurred in the high-power group.

Conclusions: PVI is feasible with HPSD PPRF in most patients using shorter total RF times. This approach appears associated with very low incidence esophageal damage than the conventional one, especially when 60W/70W and shorter application time are used.

Total RF time

The free consultation period for this content is over.

It is now only available year-round to ESC Professional Members, Fellows of the ESC, and Young combined Members

Get your access to resources

Join now
  • 1ESC Professional Members – access all ESC Congress resources 
  • 2ESC Association Members (Ivory, Silver, Gold) – access your Association’s resources
  • 3Under 40 or in training - with a Combined Membership, access all resources
Join now

Our sponsors

ESC 365 is supported by Bayer, Boehringer Ingelheim and Lilly Alliance, Bristol-Myers Squibb and Pfizer Alliance, Novartis Pharma AG and Vifor Pharma in the form of educational grants. The sponsors were not involved in the development of this platform and had no influence on its content.

logo esc

Our mission: To reduce the burden of cardiovascular disease

Who we are